MEASUREMENT MODES
60
ple.
EFM cantilevers cannot be used, due to their insufficient width.
- Install the cantilever as described in section Installing the cantilever (p.28).
- Select the Spreading resistance mode in the Operating mode panel.
- Open the Z-Controller panel.
- Verify the force Set point.
- Set the Tip voltage in the section Tip Properties.
To display the measurement results:
- Create a new chart in the Imaging window by clicking in the Chart
bar.
- If necessary, increase the size of the window to make space for the new
chart.
- Select the ‘Color map’ chart type by .
- Select the ‘Tip current’ signal by .